Count rate performances of TTVO3: the CEA-LETI time-of-flight positron emission tomograph

机译:TTVO3的计数率性能:CEA-LETI飞行时间正电子发射断层扫描

获取原文
获取原文并翻译 | 示例
           

摘要

The authors present the count rate performance of the CEA-LETI TTVO3 time-of-flight positron emission tomography (PET) system using both physical measurements and H/sub 2//sup 15/O bolus human myocardial studies. They also present a comparison between the counting statistics of H/sub 2//sup 15/O brain studies performed on this machine and on the latest available high-resolution brain bismuth germanate (BGO) tomograph, the ECAT 953B/31. During the 80 mCi cerebral blood flow study, the count rate reached 100 K events/s, and the same experiment performed on a high-resolution BGO brain machine gave only a 30% increase in signal. These results demonstrate that TTVO3 is particularly suitable for H/sub 2//sup 15/O flow studies.
机译:作者通过物理测量和H / sub 2 // sup 15 / O大剂量人体心肌研究,介绍了CEA-LETI TTVO3飞行时间正电子发射断层扫描(PET)系统的计数率性能。他们还介绍了在此机器上进行的H / sub 2 // sup 15 / O脑研究的计数统计数据与最新可用的高分辨率德国锗酸铋(BGO)断层扫描仪ECAT 953B / 31之间的比较。在80 mCi脑血流量研究期间,计数率达到了100 K个事件/秒,在高分辨率BGO脑机上进行的相同实验仅使信号增加了30%。这些结果表明,TTVO3特别适用于H / sub 2 // sup 15 / O流动研究。