首页> 外文期刊>Japanese Journal of Applied Physics. Part 1, Regular Papers, Brief Communications & Review Papers >103-km-Long Remote Measurements of End Standards Using Low-Coherence Optical-Fiber Tandem Interferometer in Experimental Room
【24h】

103-km-Long Remote Measurements of End Standards Using Low-Coherence Optical-Fiber Tandem Interferometer in Experimental Room

机译:在实验室内使用低相干光纤串联干涉仪进行最终标准物的103公里长距离远程测量

获取原文
获取原文并翻译 | 示例
           

摘要

The length information of end standards is transmitted through a 103-km-long optical fiber in the experimental room. The system uses a low-coherence tandem interferometer and a single-mode optical fiber in the wavelength range of 1500 nm. Interference fringes are generated when the optical path difference of the first interferometer compensates that of the second interferometer. The end standards of up to 50 mm in length have been calibrated remotely with a standard deviation of about 34 nm.
机译:最终标准的长度信息通过实验室内的103公里长的光纤传输。该系统使用低相干串联干涉仪和波长为1500 nm的单模光纤。当第一干涉仪的光程差补偿第二干涉仪的光程差时,会产生干涉条纹。远端校准的最长标准长度为50 mm,其标准偏差约为34 nm。