Increasing lead burden correlates with externalized cables during systematic fluoroscopic screening of Riata leads

机译:在Riata引线的系统荧光检查中,引线负担增加与外部电缆相关

获取原文
获取原文并翻译 | 示例
           

摘要

Purpose Riata and Riata ST defibrillator leads (St. Jude Medical, Sylmar, CA, USA) have been recalled due to increased risk of insulation failure leading to externalized cables. As this mechanical failure does not necessarily correlate with electrical failure, it can be difficult to diagnose. Fluoroscopic screening can identify insulation failure. Studies have suggested that insulation failure is predominantly seen in 8-Fr, single-coil models. Our patients have exclusively dual-coil leads and a high proportion of 7-Fr leads.