A new DOI detector design using discrete crystal array with depth-dependent reflector patterns and single-ended readout

机译:一种新的DOI检测器设计,使用具有与深度相关的反射器图案和单端读数的离散晶体阵列

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We developed a depth of interaction (DOI) positron emission tomography (PET) detector using depth-dependent reflector patterns in a discrete crystal array. Due to the different reflector patterns at depth, light distribution was changed relative to depth. As a preliminary experiment, we measured DOI detector module crystal identification performance. The crystal consisted of a 9×9 array of 2mm×2mm×20mm lutetium-yttrium oxyorthosilicate (LYSO) crystals. The crystal array was optically coupled to a 64-channel position-sensitive photomultiplier tube with a 2 mm×2 mm anode size and an 18.1 mm×18.1 mm effective area. We obtained the flood image with an Anger-type calculation. DOI layers and 9×9 pixels were well distinguished in the obtained images. Preclinical PET scanners based on this detector design offer the prospect of high and uniform spatial resolution.
机译:我们使用离散晶体阵列中与深度相关的反射器图案开发了相互作用深度(DOI)正电子发射断层扫描(PET)检测器。由于深度上不同的反射器图案,光的分布相对于深度发生了变化。作为初步实验,我们测量了DOI检测器模块的晶体识别性能。该晶体由2mm×2mm×20mm的oxy正氧化钇钇晶体(LYSO)晶体的9×9阵列组成。晶体阵列与阳极尺寸为2mm×2mm,有效面积为18.1mm×18.1mm的64通道位置敏感光电倍增管光学耦合。我们通过Anger型计算获得了洪水图像。在获得的图像中可以很好地区分DOI层和9×9像素。基于这种探测器设计的临床前PET扫描仪提供了高且均匀的空间分辨率的前景。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号