首页> 外文期刊>Radiation Protection Dosimetry >Lead composite vs. nonlead protective garments: Which are better? A multivendor comparison
【24h】

Lead composite vs. nonlead protective garments: Which are better? A multivendor comparison

机译:含铅复合材料与无铅防护服:哪种更好?多厂商比较

获取原文
获取原文并翻译 | 示例
       

摘要

This study aims to provide more insight in attenuation characteristics and corresponding lead (Pb) equivalences of a broad range of commercially available lead composite and nonlead protective garments. Thirty garments of five manufacturers (listed as 0.25-0.35-0.50mm Pb equivalent) were tested. Transmission values were determined at 70, 90 and 110 kVp using an inverse broad beam geometry. Pb equivalence was determined using lead sheets as reference material. A substantial variability in photon transmission across garments was found. Differences between lead composite and nonlead garments were not statistically significant. Depending on tube voltage, between 9 and 12 out of 30 garments had a lower Pb equivalence than the indicated value. This work shows that lead equivalence as indicated on a garment's label may overestimate its protective performance. Depending on the application a more thorough verification of the effectiveness of protective garments at the desired kVp is warranted.
机译:这项研究旨在提供更广泛的市场了解的铅复合和非铅防护服的衰减特性和相应的铅(Pb)当量的见解。测试了五家制造商的30件服装(等效于0.25-0.35-0.50mm Pb列出)。使用反宽束几何形状在70、90和110 kVp下确定透射率值。使用铅板作为参考材料确定铅当量。发现跨服装的光子传输存在很大的差异。含铅复合材料和无铅服装之间的差异在统计学上不显着。取决于灯管电压,在30件衣服中有9到12件之间的Pb当量低于指示值。这项工作表明,服装标签上的铅当量可能高估了其防护性能。根据不同的应用,需要对所需防护电压下的防护服的有效性进行更彻底的验证。

著录项

  • 来源
    《Radiation Protection Dosimetry》 |2017年第4期|460-465|共6页
  • 作者单位

    Radboud University Medical Center, Department of Radiology and Nuclear Medicine, Geert Grooteplein-Zuid 10, Nijmegen, Netherlands;

    Leiden University Medical Center, Department of Radiology, Albinusdreef 2, Leiden, Netherlands;

    Radboud University Medical Center, Department of Radiology and Nuclear Medicine, Geert Grooteplein-Zuid 10, Nijmegen, Netherlands;

    Franciscus Gasthuis and Vlietland, Department of Radiology, Kleiweg 500, Rotterdam, Netherlands;

  • 收录信息 美国《科学引文索引》(SCI);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号