摘要

The intercomparison measurements with Tn-Rn discriminative monitor were conducted at the Radon Laboratory of Nanhua University. The intercomparison results indicates that the Tn-Rn discrimination of most monitors (including the solid-state nuclear track detector of participant A and participant B, continuous monitors of RAD7 and RTM2100) keeps well in a wide range of the ratio of Tn to Rn. There are great differences of Tn measurement results among RAD7, RTM2100 and reference instrument. The discussion on the reason of these differences reveals the necessity of improvement on monitor calibration, Tn source make and the necessity of more international intercomparisons.
机译:用Tn-Rn判别监测仪进行比对测量是在南华大学Radon实验室进行的。相互比较的结果表明,大多数监测器(包括参与者A和参与者B的固态核径迹检测器,RAD7和RTM2100的连续监测器)的Tn-Rn区分在Tn与Rn的比率范围很广的情况下保持良好。 RAD7,RTM2100和参考仪器之间的Tn测量结果差异很大。对这些差异原因的讨论表明,有必要改进监测器校准,Tn来源制造以及进行更多国际比较的必要性。