首页> 外文期刊>Radiation Protection Dosimetry >QA PROGRAMME FOR RADON AND ITS SHORT-LIVED PROGENY MEASURING INSTRUMENTS IN NRPI PRAGUE
【24h】

QA PROGRAMME FOR RADON AND ITS SHORT-LIVED PROGENY MEASURING INSTRUMENTS IN NRPI PRAGUE

机译:NRPI布拉格的RA质量保证计划及其短寿命子代测量仪器

获取原文
获取原文并翻译 | 示例
           

摘要

To subserve the institutional research and tasks coming out from the Czech National Radon Programme, a new QA programme to calibrate all the known types of devices that measure radon and its short-lived progeny was developed at the Department of Radon mobile group of the National Radiation Protection Institute (NRPI) at Prague. The programme also included calibration of instruments measuring a unique quantity of unattached and attached fractions of short- lived radon progeny Generally, NRPI declares estimation of radon concentration during all routine calibration measurements with an overall uncertainty <5% (one sigma) and of equilibrium-equivalent radon concentration with an overall uncertainty <10% (one sigma). The results of the comparative measurements of the unattached and attached fractions of each short-lived radon progeny carried out with a comparing continuous monitor Fritra 4 in the German reference radon chamber at PTB Braunschweig indicated an acceptable level of agreement, up to 10%.
机译:为了支持捷克国家Rad计划的机构研究和任务,国家辐射的mobile部移动小组开发了一个新的质量保证计划,用于校准测量ra及其短寿命子代的所有已知类型的设备。布拉格保护研究所(NRPI)。该计划还包括校准测量短寿命ra子后代未附着和附着部分的独特数量的仪器。通常,NRPI宣布在所有常规校准测量期间估算estimation浓度,总体不确定度<5%(一个σ),并且平衡-等效ra浓度,总体不确定度<10%(一个西格玛)。在德国不伦瑞克PTB的德国参比chamber室中对连续short监视子Fritra 4进行比较,对每个短寿命ra子后代的未附着级分和附着级分的比较测量结果表明,可以接受的协议水平为10%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号