Radiation doses to patients from some common paediatric X-ray examinations were studied in three hospitals in Khartoum state, Sudan. Entrance surface dose (ESD) was determined from exposure settings using DosCal software. Totally, 459 patients were included in this study. Mean ESDs obtained from anteroposterior projection for chest, skull, abdomen and pelvis for neonates falls in the range of 52-100, 115-169, 145-183, 204-242μGy, respectively. For a 1-y-old infant, mean ESD range was 80-114, 153-202, 204-209, 181-264 μGy, respectively. Some doses for neonates and infants were exceeding the reference doses by > 20%. The results highlighted that a good technique has to adhere to guidelines necessarily. As demonstrated elsewhere, patients' doses were high in departments using single-phase generators compared with those using constant potential. The results presented will serve as a baseline data needed for deriving reference doses for paediatric X-ray examinations in Sudan.
机译:在苏丹喀土穆州的三家医院研究了一些普通的儿科X射线检查对患者的辐射剂量。使用DosCal软件根据暴露设置确定入射表面剂量(ESD)。本研究共纳入459名患者。从新生儿的胸部,颅骨,腹部和骨盆的前后投影获得的平均ESD分别在52-100、115-169、145-183、204-242μGy的范围内。对于1岁的婴儿,平均ESD范围分别为80-114、153-202、204-209、181-264μGy。新生儿和婴儿的某些剂量超出参考剂量> 20%。结果突出表明,一种好的技术必须必须遵循准则。正如其他地方所证明的,与使用恒定电位的患者相比,使用单相发电机的患者的剂量较高。所提供的结果将作为获得苏丹小儿X射线检查参考剂量所需的基线数据。