FILTERS FOR DOSE REDUCTION IN CONVENTIONAL X RAY EXAMINATIONS OF CHILDREN

机译:儿童常规X射线检查中减少剂量的过滤器

获取原文
获取原文并翻译 | 示例
           

摘要

The effects of five filtrations on dose reduction and image qualityin paediatric chest radiography with a high voltage technique are presented. The radiographs were performed using a Sieregraph unit )siemens) with 3.5 mm Al of primary total filtration, and added: 1 mm Al +0.1 mm 'Cu, 1 mm Al+0.2 mm Cu, 4 mm Al, 4 mm Al+0.1 mm Cu. The entrance air kerma (including backscatter) for each child examined was measured using TLD-100, and the parameters of expsoure were recorded. The absorbed organ doses and effective doses were calcualted using Monte Carlo simulation.
机译:提出了五种过滤对小儿胸部X光片剂量减少和图像质量的影响。使用Sieregraph单位(Siemens)进行射线照片,并进行一次总过滤3.5毫米铝,并添加:1毫米Al +0.1毫米'Cu,1毫米Al + 0.2毫米Cu,4毫米Al,4毫米Al + 0.1毫米铜使用TLD-100测量每个接受检查的儿童的入口空气比释动能(包括反向散射),并记录曝光参数。使用蒙特卡洛模拟计算吸收的器官剂量和有效剂量。