首页> 外文期刊>Radiation Protection Dosimetry >COMPARISON OF THE EUROPEAN AND NORDIC PROTOCOLS ON DOSIMETRY IN MAMMOGRAPHY INVOLVING A STANDARD PHANTOM
【24h】

COMPARISON OF THE EUROPEAN AND NORDIC PROTOCOLS ON DOSIMETRY IN MAMMOGRAPHY INVOLVING A STANDARD PHANTOM

机译:涉及标准幻象的乳房X光摄影中欧洲和北欧协议的饮食学比较

获取原文
获取原文并翻译 | 示例
           

摘要

The purpose of this work was to implement the European Protocol on Dosimetry in Mammography is Sweden. The standard average glandular dose, AGD, was measured on 32 mammography units in southern Sweden, according tio the European protocol as well as to the Nordic protocol. The most important difference is that the European protocol evaluates the half value layer, HVL, of the X ray beam with the ecompression paddle present, while it is absent in the Nordic protocol. This results in considerably higher HVL values from the European protocol and thus to higher conversion factors for the calculation of the standard AGD.
机译:这项工作的目的是实施瑞典的《欧洲X线放射剂量学议定书》。根据欧洲规程和北欧规程,在瑞典南部的32台乳腺X光摄影设备上测量了标准的平均腺体剂量AGD。最重要的区别是,欧洲协议对存在压缩板的X射线束的半值层HVL进行了评估,而北欧协议中则没有。这将导致欧洲协议的HVL值大大提高,从而导致标准AGD计算的转换因子更高。