RADIOGRAPHIC OPTICAL DENSITY UNDER AUTOMATIC EXPOSURE CONTROL AND COMPARISON WITH THE EC QUALITY CRITERIA

机译:自动曝光控制下的射线照相光学密度及其与EC质量标准的比较

获取原文
获取原文并翻译 | 示例
           

摘要

As part of a quality control programme in diagnostic radiology, automatic exposure systems in general radiography and mammography have been checked over the years 1996 to 1998. Thue optical density has been measured of films of standard phantoms exposed under normal clinicail conditions, representative of the mean density of clinical films. Approximately 100 general diagnostic systems and 28 mammographic systems, responsible for approximately 1,000,000 patient examinations, were involved.
机译:作为诊断放射学质量控制计划的一部分,在1996年至1998年期间检查了一般放射线照相和X线摄影术中的自动曝光系统。已经测量了在正常临床条件下曝光的标准体模胶片的光密度,代表了平均值临床胶片的密度。涉及大约100个一般诊断系统和28个乳房X线照相系统,负责大约1,000,000个患者检查。