A six-channel two-color interferometer has been developed for plasma electron density measurements in the C-2 field reversed configuration experiment. A CO2 laser is utilized as the main probe beams, while copropagating visible HeNe laser beams are mainly sensitive to vibration. Density measurements in C-2 plasmas have shown that this is a reliable turn-key system. The maximum residual phase noise after vibration compensation is less than ±5°, corresponding to a line integral density of 3×1018 m−2. The time resolution for routine operation is 2 μs. © 2010 American Institute of Physics Article Outline INTRODUCTION DESCRIPTION OF THE DIAGNOSTIC SYSTEM System layout Lasers Phase detection electronics Safety interlock and system operation EXPERIMENTAL DATA
机译:已经开发出一种六通道双色干涉仪,用于在C-2场反向配置实验中测量等离子体电子密度。 CO2 激光被用作主要探测光束,而可见光HeNe激光的共传播主要对振动敏感。在C-2等离子体中进行的密度测量表明,这是一个可靠的交钥匙系统。振动补偿后的最大残留相位噪声小于±5°,对应于3×1018 m−2 的线积分密度。常规操作的时间分辨率为2μs。 ©2010美国物理研究所文章大纲诊断系统说明系统布局激光器相位检测电子设备安全联锁和系统操作实验数据