We have devised a simple method for extending the zero‐field calibration of a germanium resistance thermometer to include the effects of magnetic fields up to 5 T. We describe the application of this method to the use of a germanium thermometer at liquid‐helium temperatures. We outline a similar procedure to take into account the temperature variation of the calibration of a Hall probe.