摘要

The influence of energy analyzer transmission function on measured field emission energy distributions has been evaluated by folding a Gaussian transmission function into the field emission total energy distribution. The resulting distributions at 0, 4.2, 20.3, 77, and 300 K are plotted as a function of the field‐ and work‐function‐dependent parameter d, and for the special case φ=4.40 eV, F=3.0×107 V/cm. A criterion is established for determining the width of an analyzer transmission function from distributions measured at the above temperatures. It is shown that the slope of the log of the energy distribution is changed by less than 1% when the analyzer width is changed from 0 to 100 mV.
机译:通过将高斯传递函数折合成场发射总能量分布,可以评估能量分析仪的传递函数对测得的场发射能量分布的影响。绘制了在0、4.2、20.3、77和300 K处的所得分布,它们是取决于场函数和功函数的参数d的函数,在特殊情况下φ= 4.40 eV,F = 3.0×107 V /厘米。建立了用于根据在上述温度下测得的分布确定分析仪传递函数的宽度的标准。结果表明,当分析仪的宽度从0变为100 mV时,能量分布的对数斜率变化小于1%。