Dewar System to Attain -140 to 550°C in an NMR Crosscoil Probe

机译:杜瓦系统在NMR交叉线圈探针中可达到-140至550°C

获取原文
获取原文并翻译 | 示例
           

摘要

An L‐shaped glass Dewar and associated thermometry which will provide temperatures from -140 to 550°C in a conventional NMR crosscoil nuclear induction probe is described. A particular temperature is maintained by streaming hot or cold nitrogen gas around the sample. Details on the relatively simple construction and mounting of the Dewar in the probe and of heating and cooling the nitrogen heat exchange gas are discussed.
机译:描述了一种L型玻璃杜瓦瓶和相关的测温仪,在常规NMR交叉线圈核感应探针中,它将提供-140至550°C的温度。通过使热氮气或冷氮气在样品周围流动来维持特定的温度。讨论了杜瓦瓶相对简单的结构和在探头中的安装以及加热和冷却氮气热交换气体的细节。