Changes in intraocular pressure after laser in situ keratomileusis.

机译:激光原位角膜磨镶术后眼压的变化。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To evaluate changes in intraocular pressure (IOP) measurements by Goldmann applanation tonometry after laser in situ keratomileusis (LASIK) for myopia and myopic astigmatism, and to assess the accuracy of Goldmann applanation tonometry measurements after LASIK in these eyes. METHODS: LASIK was performed on 166 eyes of 93 patients for correction of myopia and myopic astigmatism. Intraocular pressure was measured by Goldmann applanation tonometry at the central and temporal parts of the cornea before and at 1, 3, 6, and 12 months after LASIK. The amount of change in IOP was computed and its relation to different variables was evaluated by regression analysis. RESULTS: Intraocular pressure measured at the center of the cornea was reduced by a mean of 3.69 +/- 1.63 mmHg after LASIK. Multiple regression analysis showed that the decrease in IOP was related to the preoperative IOP and the change in central corneal thickness after LASIK. Measurements of IOP at the temporal part of the cornea were also reduced by a mean of 2.39 +/- 1.71 mmHg. There was wide variability in the amount of difference between the temporal and central measurements after LASIK (temporal measurements were higher than central by 0 to +4 mmHg). CONCLUSION: LASIK for myopia produced underestimation of IOP measured by Goldmann applanation tonometry at the central part of the cornea by a mean of 3.69 +/- 1.63 mmHg. The decrease of IOP was related to preoperative IOP and the change in central corneal thickness after LASIK. Temporal Goldmann applanation tonometry measurements, although decreased after LASIK, were less reliable.
机译:目的:评估通过激光原位角膜磨镶术(LASIK)治疗近视和近视散光后通过戈德曼压平眼压法测量眼内压(IOP)的变化,并评估在这些眼睛中进行LASIK手术后戈德曼压平眼压法测量的准确性。方法:对93例患者的166只眼进行LASIK手术,以矫正近视和近视散光。通过Goldmann压平眼压计在LASIK术前和术后1、3、6和12个月在角膜的中央和颞部测量眼内压。计算IOP的变化量,并通过回归分析评估其与不同变量的关系。结果:LASIK手术后,在角膜中央测得的眼内压平均降低了3.69 +/- 1.63 mmHg。多元回归分析表明,LASIK术后眼压降低与术前眼压及角膜中央厚度的变化有关。角膜颞部IOP的测量值也平均降低了2.39 +/- 1.71 mmHg。 LASIK手术后,时间测量值和中心测量值之间的差异存在很大差异(时间测量值比中心测量值高0至+4 mmHg)。结论:通过戈德曼压平眼压计测得的近视眼LASIK低估了眼压,平均眼压为3.69 +/- 1.63 mmHg。眼压的降低与术前眼压和LASIK术后角膜中央厚度的变化有关。时间性Goldmann压平眼压测量,尽管在LASIK手术后下降,但可靠性较差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号