Practical test of a point-source integrating cavity absorption meter: the performance of different collector assemblies

机译:点源积分腔吸收仪的实际测试:不同收集器组件的性能

获取原文
获取原文并翻译 | 示例
           

摘要

A prototype point-source integrating cavity absorption meter (PSICAM) is presented and compared with spectrophotometric absorption measurements. Different light collector assemblies of the PSICAM were tested regarding their capability to determine the absorption of water constituents accurately over a wide range of concentrations and scattering properties. The PSICAM setup with a radiance-type sensor showed the best performance. It was compared with a photometric absorption determination using nonscattering dye solutions. The mean difference between both methods was less than 2.4percent in the spectral range of 400-700 nm. The absorption determination with the PSICAM, when equipped with a radiance sensor as a light collector, was only little affected by scattering and temperature. We conclude that the PSICAM can be used to determine the absorption of natural seawater samples at ambient temperatures.
机译:提出了一种原型点源积分腔吸收仪(PSICAM),并将其与分光光度吸收法进行了比较。测试了PSICAM的不同集光器组件在宽范围的浓度和散射特性下准确确定水成分吸收的能力。带有辐射型传感器的PSICAM设置显示出最佳性能。将其与使用非散射染料溶液的光度吸收测定法进行了比较。两种方法之间的平均差在400-700 nm的光谱范围内小于2.4%。当配备有辐射传感器作为光收集器时,使用PSICAM进行的吸收测定几乎不受散射和温度的影响。我们得出的结论是,PSICAM可用于确定环境温度下天然海水样品的吸收率。