BACKGROUND AND OBJECTIVES: To validate accuracy and reproducibility of the Perkins tonometer, pneumatonometer, and Tono-Pen XL (Medtronic Solan, Jacksonville, FL) in estimating intraocular pressure (IOP) in rabbits. MATERIALS AND METHODS: IOP was increased from 5 to 50 mm Hg in 5-mm increments. Measurements were compared to readings of two digital manometers simultaneously measuring real IOP in the anterior chamber and vitreous cavity. Interobserver accuracy was evaluated using 4 eyes with the Perkins tonometer. RESULTS: The Perkins tonometer and Tono-Pen XL underestimated IOP and were more accurate at pressures less than 30 mm Hg. No statistically significant difference was found between real IOP and Tono-Pen XL readings. The pneumatonometer overestimated pressures in the low ranges but was accurate at pressures greater than 40 mm Hg. The Tono-Pen XL had more variability than the Perkins tonometer and pneumatonometer at high IOP. CONCLUSIONS: None of the tonometers are accurate or reproducible in estimating IOP in rabbits over the tested range. Pneumatonometry, although not very accurate, has the advantage of having acceptable variability.
机译:背景和目的:验证精度和帕金斯眼压计的再现性,pneumatonometer, Tono-Pen XL(美敦力公司塘鹅,杰克逊维尔,FL)的估计眼内压(IOP)的兔子。和方法:眼压从5提高到50毫米汞柱在5毫米的增量。同时阅读两个数字压力计测量实际前房和眼压玻璃腔。评估使用4与帕金斯的眼睛眼压计。Tono-Pen XL低估了眼压和更多准确的压力小于30毫米汞柱。不统计上的显著差异之间真正的眼压和Tono-Pen XL读数。pneumatonometer高估了压力低范围,但准确的压力更大比40毫米汞柱。Tono-Pen XL比帕金斯眼压计和可变性pneumatonometer在高眼压。眼压计是准确或可再生的估计在兔子眼压检测范围。Pneumatonometry,虽然不是很准确,的优点有可以接受的差异性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号