首页> 外文期刊>Acta crystallographica.Section D. Biological crystallography >Improving diffraction by humidity control: A novel device compatible with X-ray beamlines
【24h】

Improving diffraction by humidity control: A novel device compatible with X-ray beamlines

机译:提高衍射湿度控制:一部小说设备与x射线beamlines兼容

获取原文
获取原文并翻译 | 示例
           

摘要

Dehydration of protein crystals is rarely used, despite being a post-crystallization method that is useful for the improvement of crystal diffraction properties, as it is difficult to reproduce and monitor. A novel device for hydration control of macromolecular crystals in a standard data-collection environment has been developed. The device delivers an air stream of precise relative humidity that can be used to alter the amount of water in macromolecular crystals. The device can be rapidly installed and is fully compatible with most standard synchrotron X-ray beamlines. Samples are mounted in cryoloops and the progress of dehydration can be monitored both optically and by the acquisition of diffraction images. Once the optimal hydration level has been obtained, cryocooling is easy to achieve by hand or by using a sample changer. The device has been thoroughly tested on several ESRF beamlines and is available to users.
机译:脱水的蛋白质晶体很少使用,尽管post-crystallization方法有助于改善晶体吗衍射性质,很难复制和监控。水合作用控制的大分子晶体标准收集环境发展。精确的相对湿度,可用于改变水的大分子晶体。与大多数标准完全兼容吗同步加速器x射线beamlines。在cryoloops和脱水的进步被监控光和的收购衍射图像。获得最佳的水合作用水平,或手工cryocooling很容易实现使用一个示例变换器。几个ESRF beamlines和全面测试对用户来说是可用的。