首页> 外文期刊>Optical Engineering >X-ray spectra from a cerium target and their application to cone beam K-edge angiography
【24h】

X-ray spectra from a cerium target and their application to cone beam K-edge angiography

机译:X-ray spectra from a cerium target and their application to cone beam K-edge angiography

获取原文
获取原文并翻译 | 示例
           

摘要

The cerium-target x-ray tube is useful for performing cone beam K-edge angiography, because K-series characteristic x-rays from the cerium target are absorbed effectively by iodine-based contrast media. The x-ray generator consists of a main controller and a unit with a high-voltage circuit and a fixed anode x-ray tube. The tube is a glass-enclosed diode with a cerium target and a 0.5-mm-thick beryllium window. The maximum tube voltage and current are 65 kV and 0.4 mA, respectively, and the focal-spot sizes are 1.3 X 0.9 mm. Cerium K-series characteristic x-rays are left, using a 3.0-mm-thick aluminum filter, and the x-ray intensity is 19.9 (mu)Gy/s at 1.0 m from the source with a tube voltage of 60 kV and a current of 0.40 mA. Angiography is performed with a computed radiography system using iodine-based microspheres 15 (mu)m in diameter. In angiography of nonliving animals, we observe fine blood vessels of approximately 100 (mu)m with high contrasts.