首页> 外文期刊>journal of magnetics >PROPELLER (periodically rotated overlapping parallel lines) Comparable Image Quality to the Conventional Cartesian Magnetic Resonance Images: A Phantom Study
【24h】

PROPELLER (periodically rotated overlapping parallel lines) Comparable Image Quality to the Conventional Cartesian Magnetic Resonance Images: A Phantom Study

机译:PROPELLER (periodically rotated overlapping parallel lines) Comparable Image Quality to the Conventional Cartesian Magnetic Resonance Images: A Phantom Study

获取原文
获取原文并翻译 | 示例
           

摘要

To the best of our knowledge, studies comparing PROPELLER and the conventional Cartesian MM have mostly been reported on motion sensitivity, and there has been no focused study on the standard phantom imaging to compare them. The purpose of this study was to investigate the effect of PROPELLER method on ACR guideline QA protocol compared to Cartesian method. In this study, we compared the ACR phantom image quality of T2WI acquired on PROPELLER and conventional Cartesian method. In addition, we performed comparisons using different AFs (acceleration factors). For quantitative analysis SNR, geometric accuracy, high-contrast spatial resolution, slice thickness accuracy, slice position accuracy and low-contrast object detectability were measured on the acquired images. There were no statistically significant differences in the 6 quantification values for both methods (p > 0.05). In conclusion, we found that PROPELLER method provided a comparable SNR to Cartesian method, regardless of the various Afs