Monte Carlo and water calorimetric determination of kilovoltage beam radiotherapy ionization chamber correction factors

机译:Monte Carlo and water calorimetric determination of kilovoltage beam radiotherapy ionization chamber correction factors

获取原文
获取原文并翻译 | 示例
           

摘要

The in-phantom calibration method for radiotherapy kilovoltage x-ray beams requires ionization chamber correction factors. The overall ionization chamber correction factor accounts for changes in the chamber response due to the displacement of water by the chamber cavity and wall, the presence of the stem and the change in incident photon energy and angular distribution in the phantom to that in air. A waterproof sheath, if required, is accounted for in a sheath correction factor. The aim of this study is to determine chamber correction factors through Monte Carlo (MC) simulations and water calorimetry measurements. Correction factors are determined for the PTW TM30013, NE2571, IBA FC65-G, IBA FC65-P and Exradin A12 ionization chambers. They are compared to experimental values obtained at the German national metrology institute Physikalisch-Technische Bundesanstalt (PTB) with their water calorimetry-based absorbed dose to water primary standard and at other national metrological institutes. An uncertainty analysis considers the contributions to the uncertainty on the chamber correction factors from the field size, photon cross sections, photon fluence spectra and chamber wall and central electrode dimensions. The MC calculated chamber correction factors are within 2.2% of unity with a standard uncertainty of 0.3%. For the 50 kV and 100 - 140 kV radiation beam qualities, the calculated correction factors deviate from the measured correction factors (with a standard uncertainty of 1%) by up to 2.6%. The calculated chamber correction factors for the PTW TM30013 and Exradin A12 are consistent with those derived from the BIPM kilovoltage primary standard. The inconsistencies between the calculated and experimental chamber correction factors indicate the need to further investigate the accuracy of kilovoltage absorbed dose to water primary standards and the use of MC simulations to determine kilovoltage beam chamber correction factors.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号