Gonorrhea is a sexually transmitted bacterial infection caused by Neisseria gonorrhoeae. Nucleic acid amplification testing is the preferred method for routine diagnosis of gonorrhea from urogenital specimens, but culture is commonly used for diagnosis of disseminated infections, including gonococcal arthritis. The Hologic Aptima Combo 2 (AC2), a transcription-mediated amplification assay, is FDA and Health Canada licensed for detection of N. gonorrhoeae and Chlamydia trachomatis from urogenital, rectal, and pharyngeal specimens, but not joint fluid. In the current study, we compared the performance of microscopy, culture, and the AC2 for detection of N. gonorrhoeae from 170 joint fluid specimens. A total of five specimens were culture-positive, whereas 14 were AC2-positive. Gram-negative diplococci, characteristic of Neisseria, were observed in only two joint fluid specimens. Complementary testing confirmed the presence of N. gonorrhoeae in seven discordant (i.e., culture-negative/AC2-positive) specimens. These results indicate that the AC2 is more sensitive than culture for the diagnosis of gonococcal arthritis.