首页> 外文期刊>光学 >正確な紫外発光ダイオードの評価に向けたスぺクトルミスマッチング補正の重要性を検討した紫外放射照度計の比較検証
【24h】

正確な紫外発光ダイオードの評価に向けたスぺクトルミスマッチング補正の重要性を検討した紫外放射照度計の比較検証

机译:正確な紫外発光ダイオードの評価に向けたスぺクトルミスマッチング補正の重要性を検討した紫外放射照度計の比較検証

获取原文
获取原文并翻译 | 示例
           

摘要

In recent years, conventional mercury lamps used for ultraviolet(UV)curing and other purposes are being replaced by UV-LEDs, which provide several important benefits such as longer lifetime, lower energy consumption, and no hazardous waste. However, difficulties of irradiance measurement of UVLED and calibration problems of UV-irradiance meters have also arisen. This study compared commercial UV-irradiance meters of seven types to investigate irradiance measurements of UV-LEDs. As reference values for comparison, irradiance values of three types of standard LEDs for UV-LED were used, and they were calibrated against the spectral responsivity standard. Comparison results demonstrated that most of the reading of UV-irradiance meters differed from our UV-LED calibration values by about 10-30 . These differences were resulted from disagreement of spectra between the reference source used in each UV-irradiance meter calibration and the UV-LED. It was improved by spectrum mismatch correction.