Demonstration tests of the alignment of Fresnel zone plate focusing optics using a full-field x-ray microscope and microbeam x-ray diffraction measurements combined with the full-field x-ray microscope were performed. It was confirmed that the full-field x-ray microscope enables direct two-dimensional observation of a microbeam with sub-micrometer spatial resolution. This allowed visualization of the misalignment of the focusing optics, resulting in accurate alignment of the optics within a short time. In addition, the microscope could be used to observe the sample as well as the microbeam, which enabled clarification of the position and two-dimensional shape of the microbeam on the sample. This realized a measurement procedure that a 100-mu m-size sample was imaged with sub-micrometer spatial resolution, and then, microbeamuse measurements were performed for only the region of interest determined by the microscope, which has been difficult with conventional microbeam applications. The combination of observations by a full-field x-ray microscope and measurements using a microbeam is expected to open a new style of measurement.