首页> 外文期刊>review of scientific instruments >Microwave Exposure System for Biological Specimens
【24h】

Microwave Exposure System for Biological Specimens

机译:Microwave Exposure System for Biological Specimens

获取原文
   

获取外文期刊封面封底 >>

       

摘要

A microwave exposure system for use in the study of the bioeffects of cw microwave radiation has been designed, built, and calibrated. The specifications to which the system was designed are as follows: single frequency of 2450 MHz; power density range at the biological sample holder of 0.01ndash;200 mWsol;cm2; uniform field at the sample holder as defined by a 10percnt; variation in power density over a 15 cm diameter and halfhyphen;power variation over a 30 cm diameter; minimum cost consistent with safety, reasonable convenience of operation, and meeting technical criteria; and total space requirement not to exceed a floor space of 4.1times;4.9 m. The assembled system costs approximately $13 000 and provides better field characteristics than those specified above.

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信