首页> 外文期刊>electroanalysis >Development of a hypoxanthine biosensor based on immobilized xanthine oxidase chemically modified electrode
【24h】

Development of a hypoxanthine biosensor based on immobilized xanthine oxidase chemically modified electrode

机译:Development of a hypoxanthine biosensor based on immobilized xanthine oxidase chemically modified electrode

获取原文
           

摘要

AbstractAn amperometric biosensor for hypoxanthine determination has been developed. The sensor uses a Nafion‐paraquat chemically modified glassy‐carbon electrode. It detects the oxygen consumed by the enzymatic reaction catalyzed by xanthine oxidase which is immobilized on the electrode surface. The sensor responds linearly to hypoxanthine over the concentration range of 1 × 10−6M −2 × 10−4M. The biosensor can be reused for more than 100 times without significant deterioration in performance. After 32 days storage at 3–5°C, the sensor response remains at 68 of the initial level. The high sensitivity, selectivity and stability of this biosensor demonstrates its practical applicability for a simple, rapid and economical determination of hypoxanthine i

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有