首页> 外文期刊>electroanalysis >Comparison of detector cell configurations in flow‐injection potentiometry
【24h】

Comparison of detector cell configurations in flow‐injection potentiometry

机译:Comparison of detector cell configurations in flow‐injection potentiometry

获取原文
           

摘要

AbstractA comparative assessment has been made of three detector cell configurations, flow‐through, wall jet, and edge jet, based upon the flow‐injection performance of a solid‐state iodide‐selective electrode. Several aspects of the observed behavior have been explained with reference to flow patterns visualized by intercalation of Eriochrome T indicator into the flow stream and filmed under high magnification. The edge jet, a thin‐layer design incorporating an inlet channel at 45° to the sensing surface of the electrode, performed well over a considerable range of cell volumes, while the wall‐jet and flow‐through arrangements were prone to flow instability at higher volumes, resulting in irregular peak profiles and irreproducible baseline recovery times. Best response and recovery were recorded for all three cells when the distance between the inlet nozzle and the electrode membrane did not

著录项

获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号