首页> 外文期刊>measurement science and technology >The effect of the phase distribution on the weight function of an electromagnetic flow meter in 2D and in the annular domain
【24h】

The effect of the phase distribution on the weight function of an electromagnetic flow meter in 2D and in the annular domain

机译:The effect of the phase distribution on the weight function of an electromagnetic flow meter in 2D and in the annular domain

获取原文
           

摘要

The weight function of multi-pair electrodes and uniform-field electromagnetic flow meters in an eccentric annular domain is obtained by conformal mapping and series expansion. Two cases are studied, one of which concerns one pair of electrodes and the other concerns two pairs of electrodes with optimum separation designed for single-phase operation. The non-uniformity of the weight function is calculated for various radii, eccentricities and the angles between the symmetrical lines of the electrodes and of the insulating phase. The results are arranged in the form of initial values with tolerances. Some typical distributions of the weight function are given graphically.

著录项

  • 来源
    《measurement science and technology》 |1997年第11期|1285-1288|共页
  • 作者

    Xiao-Zhang Zhang;

  • 作者单位

    Department of Engineering Physics, Tsinghua University, Beijing, 100084, People's Republic of China;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 英语
  • 中图分类
  • 关键词

获取原文