Further measurements of the temperature dependence of the voltage response of piezoelectric tube scanner PZThyphen;5A for use in scanning tunneling microscopes
展开▼
机译:Further measurements of the temperature dependence of the voltage response of piezoelectric tube scanner PZThyphen;5A for use in scanning tunneling microscopes
We report further measurements of the temperature dependence of the piezotube coefficient for material PZThyphen;5A. These new results support our earlier findings that the temperature dependence of the piezoresponse is linear to within 10percnt; and is reduced by a factor of 5.5 on decreasing temperature from 300 to 4.2 K. Our new data imply that the piezocoefficient temperature dependence is reproducible for tubes of various sizes, even when the room temperature coefficients differ by more than 20percnt; from their published values. copy;1994 American Institute of Physics.