Comparison of a mosaic-crystal spectrometer to a high-performance solid- state detector for x-ray microfluorescence analysis

机译:马赛克晶体光谱仪与用于X射线微荧光分析的高性能固态检测器的比较

获取原文

摘要

The minimum-detectable-limit of a compact double-focusing graphite mosaic- crystal spectrometer is compared to the minimum-detectable-limit from a high- performance Ge solid-state detector. The solid angle and efficiency of the solid-state detector is much greater than for the crystal spectrometer. However, the better signal-to-noise of the spectrometer and its insensitivity to matrix fluorescence and scattering can give it a better minimum-detectable-limit for trace element analysis. The relative advantages of the two detectors are illustrated for some simple test samples. The performance of the crystal spectrometer compared to the solid-state detector increases as the flux in the x-ray probe increases. This makes crystal spectrometers especially interesting for use with new high intensity 3rd generation synchrotron microprobes. An estimate is made of the source and sample conditions favored for each detector.