A useful polarographic oxygen tension (PO2) sensing electrode system has been constructed by the incorporation of a sensing electrode (cathode) in the wall of an intravascular catheter. The reference electrode (anode) is located in the Luer hub of the catheter remote from the sensing electrode. Electrical continuity between the cathode and anode is maintained by an electrolyte solution contained within the central lumen of the catheter. The device can accurately and continuously monitor changes in oxygen tension. Use of the sensor in a limited number of adults and newborn infants indicates that it reliably and accurately measures intravascular oxygen tension.