PERANCANGAN DAN PEMBUATAN MODUL ECG DAN EMG DALAM SATU UNIT PC Sub Judul : PEMBUATAN RANGKAIAN ECG DAN PERANGKAT LUNAK ECG PADA PC

机译:一台pC机上心电图和肌电图模块的设计和制造子标题:在pC上制作心电图构建和心电图软件

摘要

Due to limited module biopac MP30 and there is no substitute modules used in the electronics lab Medika, making inefficient implementation Medika electronics lab. So that created an ECG and EMG module with output display on a PC. In making this ECG module required an instrument amplifier with 5000 times reinforcement, notch filter with cutoff frequency of 50 Hz to filter noise caused by the grid PLN, Low Pass Filter with cutoff frequency of 40 Hz to filter noise carried by strengthening signal, an ADC0804 to convert analog signals generated into a digital signal from amplifier circuit and a minimum system with a series of RS232 AT89S51 for communication with a PC. From the ECG module test results are made compared with the measurement with MP30 biopac module obtained errors in the BPM that produced by 24.7% and for the resulting amplitude R signal has an error of 19.9%. Keyword : Instrument Amplifier, Low Pass Filter, Notch Filter, Minimum System
机译:由于biopac MP30模块的数量有限,因此Medika电子实验室没有替代模块,因此Medika电子实验室的实施效率低下。这样就创建了ECG和EMG模块,并在PC上显示输出。在制造此ECG模块时,需要一个具有5000倍增强功能的仪表放大器,一个截止频率为50 Hz的陷波滤波器以滤除电网PLN引起的噪声,一个截止频率为40 Hz的低通滤波器以过滤由增强信号携带的噪声,ADC0804使用一系列RS232 AT89S51将放大电路和最小系统产生的模拟信号转换为数字信号,以便与PC进行通信。从ECG模块获得的测试结果与MP30 biopac模块的测量结果相比较,得出的BPM误差为24.7%,并且所产生的幅度R信号的误差为19.9%。关键字:仪表放大器,低通滤波器,陷波滤波器,最小系统

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号