Abstract: A high resolution and high sensitivity spectroscopic technique is developed by utilizing a Bruker IFS 120HR Fourier-transform spectrometer to detect the absorption spectra of samples inside a Ti:Sapphire laser cavity. A trigger-sampling system makes it possible to investigate the evolution of intra-cavity laser absorption spectra with the generation time. Owing to the multiplex advantage of FTS, the quantum noise in ICLAS can be significantly reduced in FT-ICLAS. The absorption of atmospheric water vapor in the range of 12450 - 12700 cm$+$MIN@1$/ was recorded at a resolution of 0.05 cm$+$MIN@1$/ to examine the performance of our FT-ICLAS apparatus. With 200 scans on the FT-ICLAS within 2 hours, a sensitivity of 10$+$MIN@9$/ cm$+$MIN@1$/ was obtained. It is also demonstrated that the FT-ICLAS is able to record both positions and intensities in a wide spectral range. This technique was applied to measure the HOD v$-OD$/ $EQ 5 overtone in the range of 12550 - 12900 cm$+$MIN@1$/ with a resolution of 0.05 cm$+$MIN@1$/. The spectrum was analyzed with the ground states combination differences method and the spectroscopic parameters were obtained. For 81.1% rotational levels in this band, the difference between calculate and experimental energy was less than 0.01 cm$+$MIN@1$/ which indicated that both the recording and analyzing of this spectrum were successful.!4
机译:摘要:利用布鲁克IFS 120HR傅里叶变换光谱仪开发了一种高分辨率,高灵敏度的光谱技术,用于检测Ti:Sapphire激光腔内样品的吸收光谱。触发采样系统使研究腔内激光吸收光谱随时间的演变成为可能。由于FTS的多重优势,可以显着降低FT-ICLAS中ICLAS中的量子噪声。以0.05 cm ++ MIN @ 1 $ /的分辨率记录了在12450-12700 cm ++ MIN @ 1 $ /范围内的大气水蒸气的吸收,以检查FT-ICLAS仪器的性能。在2小时内用FT-ICLAS进行了200次扫描,灵敏度为10 $ + MIN @ 9 $ / cm $ + MIN @ 1 $ /。还证明了FT-ICLAS能够在很宽的光谱范围内记录位置和强度。该技术用于测量HOD v $ -OD $ / $ EQ 5泛音,范围为12550-12900 cm $ + $ MIN @ 1 $ /,分辨率为0.05 cm $ + $ MIN @ 1 $ /。用基态组合差法分析光谱,得到光谱参数。对于该频带中81.1%的旋转水平,计算能量与实验能量之差小于0.01 cm $ + $ MIN @ 1 $ /,这表明该光谱的记录和分析均成功完成了!! 4

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号