首页> 外文会议>4th International Symposium on Bioelectronics and Bioinformatics >PD of two kinds of blood pulse wave as an alternative of PAT
【24h】

PD of two kinds of blood pulse wave as an alternative of PAT

机译:两种脉搏波的PD替代PAT

获取原文
获取原文并翻译 | 示例

摘要

Peak Delay (PD) of two different blood pulse waves measured at the same site is intensively studied in this paper. The derived expression, based on Womersley theory and simplified Navier-Stokes equations, implies that PD is closely related to the change of the Young's modulus of the blood vessel, which is promising in noninvasively BP monitoring. Traditionally, BP is estimated with Pulse Transit Time (PTT) or Pulse Arrival Time (PAT). Experiment results show that PAT and PD are inversely related with correlation coefficient from 0.78 to 0.95 when BP is relatively stable. If BP changes remarkably, calibration should be conducted to better trace PAT with PD. Therefore PTT or PAT can be replaced by PD for continuous and noninvasive BP monitoring, with the convenience that the measuring site is concentrated on fingertip.
机译:本文深入研究了在同一位置测量的两种不同脉搏波的峰延迟(PD)。基于Womersley理论和简化的Navier-Stokes方程得出的表达式表示PD与血管的杨氏模量的变化密切相关,这在无创BP监测中很有希望。传统上,BP是用脉冲传输时间(PTT)或脉冲到达时间(PAT)估算的。实验结果表明,当BP相对稳定时,PAT和PD与相关系数从0.78到0.95成反比。如果BP发生显着变化,则应进行校准以更好地跟踪PD和PD。因此,可以用PD代替PTT或PAT,以进行连续和无创的BP监测,并且方便将测量部位集中在指尖上。