【24h】

OPTIMIZATION OF ULTRASONIC TRANSDUCER POSITIONING

机译:超声传感器定位的优化

获取原文
获取原文并翻译 | 示例

摘要

Typical methods of attaching ultrasonic transmitters and receivers to the outside of a fluid flow pipe involve substantial setup and calibration time, and the necessity to quantify a variety of system variables values that may not be precisely known. A unique method of adjusting the placement of the transceivers of an ultrasonic measurement system as been developed at the Instrumentation and Cryogenics Laboratory of the University of Southern Mississippi. The positioning system was designed to analyze the signal available at the receiver and based upon the results of signal analysis, verify that the transducers are placed in locations that provide the optimum acoustical reception. Results of he positioning system were compared with results from two off-the-shelf ultrasonic flow meters and with analytical calculations. Future design modifications include adapting the measurement configuration to cryogenic systems.
机译:将超声发射器和接收器连接到流体流管外部的典型方法涉及大量的设置和校准时间,以及量化各种可能无法精确得知的系统变量值的必要性。南密西西比大学的仪器与低温实验室已经开发出一种独特的调整超声测量系统收发器位置的方法。定位系统旨在分析接收器处可用的信号,并基于信号分析的结果,验证将换能器放置在提供最佳声学接收的位置。将定位系统的结果与两个现成的超声波流量计的结果进行比较,并进行分析计算。未来的设计修改包括使测量配置适应低温系统。