首页> 外文会议>ASMS Conference on Mass Spectrometry and Allied Topics >Integration of Paper Spray Ionization-High Field Asymmetric Waveform Ion Mobility Spectrometry-Mass Spectrometry (PSI-FAIMS-MS) for Forensic Applications
【24h】

Integration of Paper Spray Ionization-High Field Asymmetric Waveform Ion Mobility Spectrometry-Mass Spectrometry (PSI-FAIMS-MS) for Forensic Applications

机译:纸张喷射电离 - 高场非对称波形离子迁移率光谱法的整合 - 质谱(PSI-FAIMS-MS)用于法医应用

获取原文

摘要

In this work PSI-FAIMS-MS analyses of explosives and drugs in negative mode and positive mode, respectively, were achieved. Baseline separation of the two groups of analytes was observed. A slightly lower curtain plate voltage (500 V) was needed to keep adduct ions intact while transmitting from PSI to FAIMS. Solvent vapor introduced as curtain gas was also needed to assist ions in going through FAIMS cell under the influence of the waveform. Although characteristic CW ions were observed in PSI-MS mode and also with the FAIMS cell in place, obtaining a CV-gram with adequate peak shape will require further experiments.
机译:在这项工作中,分别实现了Psi-Faims-MS分析负模式和积极模式的爆炸物和药物分析。观察到两组分析物的基线分离。需要稍低的窗帘电压(500V)来保持加合离子在从PSI传输到FAIMS时完整。作为帘气体引入的溶剂蒸气也需要在波形的影响下辅助离子通过Faims Cell。尽管在PSI-MS模式下观察到特征CW离子,并且在配对电池到位,但以足够的峰形状获得CV-GRAM将需要进一步的实验。