Oxygen and moisture sensitive complexes like metallocene catalysts were analyzed using APCI and APPI ionization sources. Two introduction methods are presented: one with a direct introduction into the mass spectrometer and one with a glove box connected to the mass spectrometer. Degradation products are observed with a higher ionization yield with APPI thanks to toluene acting as dopant. CID fragments in gas phase which react with trace of water in the mass spectrometer are only observed with APCI. The first analysis of metallocene by ion mobility mass spectrometry was shown. Isomeric information can be obtained for this type of catalyst.
机译:使用APCI和APPI电离来源分析氧气和水分敏感复合物等茂金属催化剂。提出了两种介绍方法:一种具有直接引入质谱仪的一个,一个带连接到质谱仪的手套箱。由于甲苯作用为掺杂剂,观察到具有更高的离子化产量的降解产物。在质谱仪中与痕量水反应的气相中的CID片段仅用APCI观察。示出了离子迁移率质谱法的第一次分析茂金属。可以获得这种类型的催化剂可以获得异构信息。

著录项