Inter-laboratory Comparison of Reflection and Transmission Measurements in WR-06 waveguide (110 GHz to 170 GHz)

机译:WR-06波导反射和透射测量的实验室间比较(110GHz到170 GHz)

获取原文

摘要

This paper describes an exercise that was undertaken recently to compare reflection and transmission measurements in rectangular metallic wave guide from 110 GHz to 170 GHz (i.e. in the WR-06 wave guide size). The comparison involved making measurements on four devices fitted with 'precision' MIL-DTL-3922-67D style flanges. These devices were circulated amongst the nine organizations that chose to participate in the exercise. The comparison took place between August 2008 and September 2009. Results from the exercise are presented in graphical form along with a statistical summary showing average variability for the measurements. The authors believe this is the first time that such a comparison of measurements has been made at these frequencies. These results therefore provide a benchmark for the current state-of-the-art for measurements made in wave guide at these frequencies.
机译:本文介绍了一个锻炼,最近进行了比较从110GHz到170GHz(即WR-06波导尺寸)的矩形金属波导中的反射和传输测量。比较涉及在配备有“精密'MIL-DTL-3922-67D”法兰的四个设备上进行测量。这些设备在九个组织中循环,选择参与练习。比较于2008年8月至2009年9月之间发生。练习的结果以图形形式呈现,以及统计摘要,显示了测量的平均变异性。作者认为这是第一次在这些频率下进行了这种测量比较。因此,这些结果为当前最先进的用于在这些频率的波导中进行测量提供基准。