The aerosol penetrations of HEPA filters are defined as 0. 1 percent or less. Test aerosol concentration at filter inlet is high but the concentration at filter outlet is low for the same aerosol-measuring instrument. Light scattering photometers are widely used to measure the filter performance at not only Japan but also USA and others. In this study, we presented a method to emulate a light scattering photometer (Sibata P-5H) using an optical particles counter (Rion KC-18). This combination provided relative concentrations ranging from 0.0001 to 10000 counts for aerosol particles of 0.1 to 1 mu m in diameter. Six HEPA filters for dust respirators were tested and compared by this method and official test method.
机译:HEPA过滤器的气溶胶渗透定义为0.1%或更低。在过滤器入口处测试气溶胶浓度高,但滤波器出口处的浓度对于相同的气溶胶测量仪器效率低。光散射光度仪广泛用于测量滤波器性能,不仅是日本,而且还可以在美国和其他人身上测量过滤器性能。在该研究中,我们介绍了一种使用光学粒子计数器(RION KC-18)模拟光散射光度计(SIBATA P-5H)的方法。这种组合提供了0.0001至10000计数的相对浓度,用于均直径为0.1至1μm的气溶胶颗粒。通过该方法和官方测试方法测试和比较六种HEPA过滤器。