Evaluation of inflow air velocity of a Biological Safety Cabinet installed in BSL2 laboratory.

机译:BSL2实验室中安装生物安全柜的流入空气速度评价。

获取原文

摘要

In this inspect ion, the standard value of wind velocity was set individually in each type of Biological Safety Cabinet. Each value was originally set by each manufacturer. These values were based on results of the microbiological examinations in BSC and past accumulation data. In order to use BSC safely it is important to keep the wind velocity within the standard value. The decrease of wind velocity is caused by the clogging of HEPA filter. The maintenance plan matched to the frequency of inspection is necessary. It seems that the handling of the lowest value of standard wind velocity is important.
机译:在该检查中,在每种类型的生物安全柜中单独设定风速的标准值。每个值最初由每个制造商设置。这些值基于BSC和过去的积累数据中的微生物检查的结果。为了安全地使用BSC,重要的是保持标准值内的风速。风速降低是由HEPA过滤器的堵塞引起的。符合检查频率的维护计划是必要的。似乎标准风速的最低值的处理很重要。