首页> 外文会议>Saratov Meeting >Theoretical and experimental study on near infrared time-resolved optical diffuse tomography
【24h】

Theoretical and experimental study on near infrared time-resolved optical diffuse tomography

机译:近红外时间分辨光学漫射断层扫描的理论与实验研究

获取原文

摘要

Parts of the works of our group in the past five years on near infrared time-resolved (TR) optical tomography are summarized in this paper. The image reconstruction algorithm is based on Newton Raphson scheme with a datatype R generated from modified Generalized Pulse Spectrum Technique. Firstly, the algorithm is evaluated with simulated data from a 2-D model and the datatype R is compared with other popularly used datatypes. In the second part of the paper, the in vitro and in vivo NIR DOT imaging on a chicken leg and a human forearm, respectively are presented for evaluating both the image reconstruction algorithm and the TR measurement system. The third part of this paper is about the differential pathlength factor of human head while monitoring head activity with NIRS and applying the modified Lambert-Beer law. Benefiting from the TR system, the measured DPF maps of the three import areas of human head are presented in this paper.
机译:我们集团在过去五年内近五年的部分作品概述了近五年的近五年的光学断层扫描。图像重建算法基于具有由修改的广义脉冲谱技术产生的数据型R的牛顿Raphson方案。首先,使用来自2-D模型的模拟数据进行评估算法,并将数据类型R与其他普遍使用的数据类型进行比较。在纸张的第二部分中,分别在鸡腿和人前臂上进行体外和体内NIR点成像,用于评估图像重建算法和TR测量系统。本文的第三部分是关于人头的差分路径,同时监测NIR的头部活动并应用改良的兰伯特啤酒法。从TR系统中受益,本文提出了人体头部三个进口区域的测量DPF地图。