Ultrasound-based measurement of relative blood volume

机译:基于超声的相对血容量测量

获取原文

摘要

This study introduces an online detection technique of relative blood volume based on ultrasound, which analyzes the correlation between changes in serum protein concentration and attenuation in sound velocity and intensity. By measuring the attenuation in sound velocity and intensity, this method is employed to calculate RBV, and then evaluate dry weight of dialysis patients. In addition, this investigation also illustrates the structure and principle of detection device and provides the calculation formula. RBV values range between 15.6%∼22.9% calculated from the model equipment. This detection method possesses multiple advantages such as real time, convenient, reproducible, non-invasive, etc.
机译:本研究介绍了基于超声波的相对血容量的在线检测技术,其分析了血清蛋白质浓度变化与声速和强度衰减之间的相关性。通过测量声速和强度的衰减,使用该方法来计算RBV,然后评估透析患者的干重。此外,该研究还说明了检测装置的结构和原理,并提供计算公式。 RBV值范围从模型设备计算的15.6%〜22.9%之间。该检测方法具有多种优点,例如实时,方便,可重复,非侵入性等。