Color stability is the key goal during the process of color dissemination and replication. The results of color consistent can be influenced by many factors. In this project, we focused on the error caused by measurement instrument, i.e. spectroradiometer. According to the radiance data collected from six British Ceramic Research Association tiles under two different standard illuminants, i.e. illuminant A and illuminant CWF, the tristimulus value of different colors under different conditions of illuminants were calculated and used for computing the mean color different from mean of the six tiles. Last, the performance of stability and consistency of the spectroradiometer was evaluated. The result showed that the spectroradiometer under standard illuminant CWF outperformed than that under standard illuminant A.
机译:颜色稳定性是颜色传播和复制过程中的关键目标。颜色的结果一致可能受到许多因素的影响。在这个项目中,我们专注于由测量仪器引起的误差,即光谱仪。根据两种不同标准光源的六个英国陶瓷研究协会瓷砖收集的辐射数据,即光源A和发光体CWF,计算不同颜色在不同的照明条件下的三刺激值,并用于计算与平均不同的平均颜色六个瓷砖。最后,评估了光谱仪的稳定性和一致性的性能。结果表明,标准光源CWF下的光谱差管优于标准光源A.