首页> 外文会议>Modularity and tapers in total joint replacement devices >Validation of an Optical Coordinate Measuring Machine for the Measurement of Wear at the Taper Interface in Total Hip Replacement
【24h】

Validation of an Optical Coordinate Measuring Machine for the Measurement of Wear at the Taper Interface in Total Hip Replacement

机译:光学坐标测量机在全髋关节置换中用于测量锥度界面处的磨损的验证

获取原文
获取原文并翻译 | 示例

摘要

If the factors affecting taper performance are to be studied, accurate and validated volumetric values of loss are important to contextualize the importance of taper damage. This study validates the volumetric loss values from the RedLux™ optical coordinate measuring machine, the sensitivity of the technique to the point-cloud density and the repeatability of the technique. The limits of agreement (95 %) between the optical coordinate measuring machines and the gravimetric values are -0.0416 mm~3 and 0.173 mm~3. The average measured angle was assessed against a calibrated taper and shown to be within 0.0024°.
机译:如果要研究影响锥度性能的因素,那么准确而有效的损耗体积值对于确定锥度损伤的重要性至关重要。这项研究验证了RedLux™光学坐标测量机的体积损失值,该技术对点云密度的敏感性以及该技术的可重复性。光学坐标测量机与重量值之间的一致性极限(95%)为-0.0416 mm〜3和0.173 mm〜3。相对于校准的锥度评估平均测量角度,并显示在0.0024°之内。