99Tcm-DTPA SPECT肾动态显像患者周围剂量率的研究

         

摘要

目的:对行锝-99标记的二乙三胺五乙酸(99Tcm-DTPA)单光子发射计算机断层成像术(SPECT)肾动态显像患者周围剂量率进行研究,为保证患者周围公众人员辐射安全提供数据依据.方法:对109例肾动态显像患者周围剂量率进行研究,在患者注射放射性药物后的不同时间,在距患者不同距离处测量剂量率,分析剂量率随时间和距离的变化.结果:在肾动态显像患者注射放射性药物后0.5 h、1 h、2 h、3 h、6 h和7 h时,从距患者4 m到0.5 m处的剂量率分别为:0.76~8.86μSv/h、0.68~7.27μSv/h、0.57~4.52μSv/h、0.56~3.90μSv/h、0.23~2.07μSv/h和0.21~1.05μSv/h,注射后24 h时,距患者各距离处的剂量率均为本底.结论:肾动态显像患者周围剂量率会随时间和距离快速降低;当注射放射性药物24 h以后,患者周围的剂量率降为本底.%Objective:To study the radiation dose rate around patients underwent renal dynamic imaging of SPECT with 99Tcm-DTPA, so as to provide the basis for ensuring the radiation safety of the surrounding environment and the general public.Methods: 109 patients who underwent renal dynamic imaging were enrolled in this study, and the radiation dose rates of different time and different distance after 99Tcm-DTPA was injected patients were measured. And then, the changes of dose rate following time and distance were analyzed. Results: At the 0.5h, 1h, 2h, 3h, 6h, and 7h after radiopharmaceuticals were injected patients underwent renal dynamic imaging, the radiation dose rates were 0.76-8.86 μSv/h, 0.68-7.27 μSv/h, 0.57-4.52 μSv/h, 0.56-3.90 μSv/h, 0.23-2.07 μSv/h and 0.21-1.05 μSv/h, respectively, between 0.5 m and 4 m away from patients. At 24h after radiopharmaceuticals were injected, all of radiation dose rates at different distances around patients were same with background values.Conclusion: The radiation dose rates around patients underwent renal dynamic imaging will rapidly decrease with the increasing of time and distance. At 24h after radiopharmaceuticals were injected, all of radiation dose rates at different distances away from patients were same with background values.

著录项

相似文献

  • 中文文献
  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号