Correlation between Low Tube Voltage in Dual Source CT Coronary Artery Imaging with Image Quality and Radiation Dose

         

摘要

The influence of low tube voltage in dual source CT(DSCT) coronary artery imaging on image quality and radiation dose and its application value in clinical practice were investigated. Totally, 300 cases of chest pain with low body mass index(BMI 0.05). The CTDIvol values in groups A, B and C were 51.35±12.21, 21.28±7.13 and 6.34±3.34 mGy, respectively, with the difference being statistically significant(P<0.05). The ED values in groups A, B and C were 9.27±1.63, 4.56±2.29 and 2.29±1.69 mSv, respectively, with the difference being statistically significant(P<0.05). It was suggested that for the patients with low BMI, the application of DSCT coronary artery imaging with low tube voltage can obtain satisfactory image quality, and simultaneously, significantly reduce the radiation dose.