An Investigation of Calibration Phantoms for CT Scanners with Tube Voltage Modulation

机译:带有管电压调制的CT扫描仪校准模型的研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The effects of calibration phantoms on the correction results of the empirical artifacts correction method (ECCU) for the case of tube modulation were investigated. To improve the validity of the ECCU method, the effect of the geometry parameter of a typical single-material calibration phantom (water calibration phantom) on the ECCU algorithm was investigated. Dual-material calibration phantoms (such as water-bone calibration phantom), geometry arrangement, and the area-ratio of dual-material calibration phantoms were also studied. Preliminary results implied that, to assure the effectiveness of the ECCU algorithm, the polychromatic projections of calibration phantoms must cover the polychromatic projection data of the scanning object. However, the projection range of a water calibration phantom is limited by the scan field of view (SFOV), thus leading to methodological limitations. A dual-material phantom of a proper size and material can overcome the limitations of a single-material phantom and achieve good correction effects.
机译:研究了校准体模对电子管调制情况下经验伪影校正方法(ECCU)校正结果的影响。为了提高ECCU方法的有效性,研究了典型的单一材料校准模型(水校准模型)的几何参数对ECCU算法的影响。还研究了双材料校准体模(如水骨校准体模),几何结构以及双材料校准体模的面积比。初步结果表明,为了确保ECCU算法的有效性,校准体模的多色投影必须覆盖扫描对象的多色投影数据。但是,水校准体模的投影范围受到扫描视场(SFOV)的限制,因此导致了方法上的限制。具有适当尺寸和材料的双材料幻影可以克服单材料幻影的局限性并获得良好的校正效果。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号