Anterior capsule contraction and intraocular lens decentration and tilt after hydrogel lens implantation

机译:水凝胶晶状体植入后前囊收缩和人工晶状体偏心和倾斜

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

AIM—To prospectively investigate changes in the area of the anterior capsule opening, and intraocular lens (IOL) decentration and tilt after implantation of a hydrogel IOL.
METHODS—100 patients underwent implantation of a hydrogel IOL in one eye and an acrylic IOL implantation in the opposite eye. The area of the anterior capsule opening, and the degree of IOL decentration and tilt were measured using the Scheimpflug videophotography system at 3 days, and at 1, 3, and 6 months postoperatively.
RESULTS—The mean anterior capsule opening area decreased significantly in both groups. At 6 months postoperatively, the area in the hydrogel group was significantly smaller than that in the acrylic group. The mean percentage of the area reduction in the hydrogel group was also significantly greater than that in the acrylic group, being 16.9% in the hydrogel group and 8.8% in the acrylic group. In contrast, IOL decentration and tilt did not progress in either group. No significant differences were found in the degree of IOL decentration and tilt throughout the follow up period.
CONCLUSIONS—Contraction of the anterior capsule opening was more extensive with the hydrogel IOL than with the acrylic IOL, but the degree of IOL decentration and tilt were similar for the two types of lenses studied.

机译:目的—前瞻性研究水凝胶人工晶状体植入后前囊开口面积的变化以及人工晶状体的偏斜和倾斜。
方法— 100例患者在一只眼睛和另一只眼中植入了水凝胶人工晶状体在另一只眼中植入丙烯酸人工晶体。在术后第3天,术后第1、3、6个月使用Scheimpflug视频摄影系统测量前囊开口的面积以及IOL偏心和倾斜的程度。
结果—前囊平均开口面积两组均显着下降。术后6个月,水凝胶组的面积明显小于丙烯酸组。水凝胶组中面积减少的平均百分比也显着大于丙烯酸基团中的平均百分比,水凝胶组中的平均减少率为16.9%,丙烯酸基团中的为8.8%。相比之下,两组的人工晶体偏心和倾斜均没有进展。在整个随访期间,IOL的偏斜和倾斜度没有发现显着差异。
结论—水凝胶IOL的前囊开口收缩比丙烯酸IOL的前囊开口更宽,但IOL的偏斜程度两种类型的镜片的倾斜度和倾斜度相似。