首页> 美国卫生研究院文献>Biomedical Optics Express >Dove prism based rotating dual beam bidirectional Doppler OCT
【2h】

Dove prism based rotating dual beam bidirectional Doppler OCT

机译:基于鸽子棱镜的旋转双光束双向多普勒OCT

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Traditional Doppler OCT is highly sensitive to motion artifacts due to the dependence on the Doppler angle. This limits its accuracy in clinical practice. To overcome this limitation, we use a bidirectional dual beam technique equipped with a novel rotating scanning scheme employing a Dove prism. The volume is probed from two distinct illumination directions with variable controlled incidence plane, allowing for reconstruction of the true flow velocity at arbitrary vessel orientations. The principle is implemented with Swept Source OCT at 1060nm with 100,000 A-Scans/s. We apply the system to resolve pulsatile retinal absolute blood velocity by performing segment scans around the optic nerve head and circumpapillary scan time series.
机译:由于对多普勒角度的依赖性,传统的多普勒OCT对运动伪影高度敏感。这限制了其在临床实践中的准确性。为了克服此限制,我们使用了双向双光束技术,该技术配备了采用道威棱镜的新型旋转扫描方案。从具有可变受控入射平面的两个不同照明方向探测体积,从而允许在任意容器方向上重建真实流速。该原理是通过1060nm扫描源OCT和100,000 A-Scans / s来实现的。我们应用该系统通过在视神经头周围进行段扫描和乳头周扫描时间序列来解决脉动性视网膜绝对血流速度。